Many critical environments would not be possible without High
Efficiency Particulate Air or HEPA Filters. HEPA filtration, invented
during the second world war, removes up to 99.999% of airborne
particulate. Without HEPA filtration today's modern clean rooms,
operating suites and other sterile environments would not be possible.
Depending upon a variety of engineering, use and environmental
factors most filters will last 2 - 10+ years. Each year the filters
should be professionally evaluated. As the filters load performance
will fall off and adjustments should be made to the mechancial
system, where possible, to compensate. Proper care improves performance
and adds to filter life.
It is common to find small leaks and/or broken seals
from year to year. We employ the ATI TDA2G Aerosol Photometer
to quickly pinpoint leaks so they can be repaired. This instrument
accurate from 0.0001% to 100%. All leaks are recorded during the
evalution, documented in the service report and, where possible,
repaired.

ATI TDA2G
A quality Laboratory Ventilation Management
program includes establishing a system to manage your HEPA filter
population. Your commissioning report is an excellent beginning.
Proper management and testing delivers dividends with peace of
mind, ease of maintenance and prolonged filter lfe.

|