Biological Safety Cabinets provide both product and personnel
protection. To ensure proper function each cabinet must be certified
at installation and annually thereafter. The recognized standard
is the National Sanitation Foundation's Standard 49 (NSF-49).
The NSF-49 certification method ensures that air balance is correct
and filters leak free. Our seasoned technical services team is
equipped with the latest equipment and training. We are proud
to provide the high quality service expected by our clients -
industry leaders who demand the best.

National Sanitation Foundation Standard (NSF-49)
PRIMARY TESTING
- Air Velocity Evaluation - Inflow and Down Flow Velocity Patterns
- HEPA Filter Challenge Leak Test - PAO
- Pressure Leak Testing
- Visible Vapor Testing - Air Flow Smoke Patterns
- Cabinet Leak Testing - Where Applicable
SECONDARY TESTING
- Electrical Safety Analysis to Include Electrical Leakage,
Ground
- Circuit Resistance and Sound Level Test
- Light Level Test
- Ultraviolet Light Level Test - Where Applicable
ADJUSTMENT AND REPAIR
- Cabinet Decontamination
- Filter Replacement
- Blower Replacement
- Complete System Rebuilding

|